Detecting radiation: the Geiger-Müller counter and background correction

Explain how a Geiger-Müller tube detects radiation, correct a count rate for background, and identify a source with absorbers.

If something looks wrong, please tell us.

We cannot see, smell or feel radiation, so we need instruments to find it. The most common one is the Geiger-Müller counter, the clicking meter seen in films and used in laboratories. This lesson explains how it works and how to turn its readings into reliable answers.

How a Geiger-Müller tube works

A Geiger-Müller (GM) tube is a sealed tube filled with a low-pressure gas, with a thin window at one end and a high voltage between a central wire and the metal wall. When radiation enters the tube, it ionises some gas atoms. The freed charges are accelerated by the high voltage and cause an avalanche of further ionisation, which gives a brief pulse of current. A counter records each pulse (or makes a click). Alpha particles are stopped by an ordinary tube wall, so a very thin (mica) end-window is used to detect them.

The count rate is the number of counts per second or per minute. It is proportional to the activity of the source, but it is only a fraction of it, since not every particle enters the tube or is detected.

Background correction

Even with no source nearby, the counter clicks from time to time because of background radiation. The count rate with the source in place is the sum of the source's count rate and the background. So: corrected count rate = measured count rate − background count rate. Always measure the background first, with the source far away, and subtract it.

Identifying radiation with absorbers

Which absorber stops what
AbsorberStopsDoes not fully stop
A sheet of paperAlphaBeta, gamma
A few millimetres of aluminiumAlpha and betaGamma
Thick leadAlpha, beta and nearly all gammaOnly very energetic gamma, which is reduced greatly

Worked examples

Example 1. A GM counter registers 300 counts in 1 minute with a source in front of it. With the source removed it registers 60 counts in 1 minute. Corrected count rate = 300 − 60 = 240 counts per minute.

Example 2 (half-life). The background is 20 counts per minute. A source gives a measured rate of 260 counts per minute at the start and 140 counts per minute 2 hours later. Corrected rates: 260 − 20 = 240 and 140 − 20 = 120. The corrected rate has halved in 2 hours, so the half-life is 2 hours. Without correcting the background, the figures 260 and 140 would not even give an exact half.

Example 3 (identifying a source). The background is 20 counts per minute. A source gives 520 counts per minute with no absorber, 520 with paper in the way, and 20 with 3 mm of aluminium. Paper makes no difference, so there is no alpha. Aluminium reduces the count to the background level, so there is no gamma either. The source emits beta particles only.

Corrected count rate = measured rate − background rate. Paper stops alpha; aluminium stops beta; thick lead is needed for gamma.

— Rule to remember

Key takeaways

  • A GM tube detects ionisation caused by radiation and counts each pulse.
  • Count rate is a fraction of the activity.
  • Always subtract the background before comparing readings.
  • Absorbers of different thickness help identify the type of radiation.

Check your understanding

  1. 1. A Geiger-Müller tube detects radiation because the radiation:

    1. Heats the gas in the tube until it expands and clicks
    2. Makes the glass of the tube glow with visible light
    3. Ionises the gas and produces a pulse of current
    4. Is changed into visible light that a lamp then records
    Show answer

    Answer: C. The radiation ionises gas atoms; the high voltage multiplies this into a pulse, which the counter records.

  2. 2. A counter reads 200 counts per minute with a source and 30 counts per minute without it. The count rate due to the source is:

    1. 230 counts per minute
    2. 170 counts per minute
    3. 200 counts per minute
    4. 30 counts per minute
    Show answer

    Answer: B. Corrected rate = 200 − 30 = 170 counts per minute.

  3. 3. The count rate from a source is not changed by a sheet of paper but falls to the background level when 3 mm of aluminium is placed in front. The source emits:

    1. Alpha particles only
    2. Gamma rays only
    3. Alpha and gamma
    4. Beta particles only
    Show answer

    Answer: D. Paper makes no difference, so no alpha is present. The fall to background with aluminium means no gamma either. That leaves beta only.

  4. 4. The corrected count rate of a source is 80 counts per minute and its half-life is 3 hours. After 6 hours the corrected count rate will be:

    1. 20 counts per minute
    2. 40 counts per minute
    3. 60 counts per minute
    4. 10 counts per minute
    Show answer

    Answer: A. 6 hours is 2 half-lives, so the rate falls to 80 ÷ 4 = 20 counts per minute.

  5. 5. Background radiation arises from:

    1. Cosmic rays, rocks and soil, and radon gas in the air
    2. Nuclear power stations only, and nowhere else
    3. Hospitals only, and not from the natural world
    4. The visible light of the Sun, and nothing else
    Show answer

    Answer: A. Background radiation comes from several natural sources, and smaller man-made ones. It is not limited to one place or one source.