- Practice
- About 16 min
- 5 practice questions
Vernier calipers and screw gauge
Find the least count of a vernier caliper and a screw gauge, take a reading, and correct it for zero error.
If something looks wrong, please tell us.
Instrument questions are a favourite in exams because they check understanding, not memory. You must know the least count, how to read the scale, and how to correct a zero error.
Least count
The least count is the smallest value an instrument can measure. For a vernier caliper: least count = value of 1 main scale division (MSD) − value of 1 vernier scale division (VSD) = 1 MSD / (number of vernier divisions). With 1 mm MSD and 10 VSD equal to 9 MSD, the least count is 0.1 mm. For a screw gauge: least count = pitch / number of circular scale divisions, where pitch is the distance moved by the screw in one full turn.
Taking a reading
- Vernier caliper: reading = main scale reading (just before the vernier zero) + (coinciding vernier division × least count).
- Screw gauge: reading = linear (main) scale reading + (circular scale reading × least count).
- Zero error: with jaws closed, a nonzero reading is a zero error. Corrected reading = observed reading − zero error. A positive zero error is subtracted; a negative zero error is added.
Worked examples
Example 1. A vernier caliper has 1 MSD = 1 mm and 50 vernier divisions equal to 49 MSD. Least count = 1 mm / 50 = 0.02 mm = 0.002 cm.
Example 2. A vernier caliper has least count 0.01 cm. The main scale reads 4.2 cm and the 6th vernier division coincides. Reading = 4.2 + 6 × 0.01 = 4.26 cm.
Example 3. A screw gauge has pitch 1 mm and 100 circular divisions, so least count = 1/100 = 0.01 mm. A wire gives a linear scale reading of 3 mm and a circular reading of 27 with a zero error of +0.03 mm. Observed = 3 + 27 × 0.01 = 3.27 mm. Corrected diameter = 3.27 − 0.03 = 3.24 mm.
Example 4. If the zero error is negative, say the zero reads 98 on the circular scale, the error is −0.02 mm. An observed reading of 2.45 mm is corrected to 2.45 + 0.02 = 2.47 mm.
Least count (screw gauge) = pitch / circular divisions. Corrected reading = observed reading − zero error (with its sign).
Key takeaways
- Least count decides the precision of an instrument.
- Always check the zero error first and apply its sign correctly.
- A smaller least count gives a smaller percentage error in the same measurement.
Check your understanding
-
1. In a vernier caliper, 1 main scale division is 1 mm and 20 vernier divisions equal 19 main scale divisions. What is the least count?
- 0.5 mm
- 0.05 mm
- 0.95 mm
- 0.02 mm
Show answer
Answer: B. Least count = 1 MSD ÷ number of vernier divisions = 1 mm ÷ 20 = 0.05 mm. (Equivalently 1 − 0.95 = 0.05 mm.)
-
2. A screw gauge has pitch 1 mm and 100 circular divisions. A wire gives a linear scale reading of 4 mm and a circular scale reading of 38, with no zero error. What is its diameter?
- 4.038 mm
- 42 mm
- 4.83 mm
- 4.38 mm
Show answer
Answer: D. Least count = 1 ÷ 100 = 0.01 mm. Reading = 4 + 38 × 0.01 = 4.38 mm.
-
3. A screw gauge has a zero error of −0.02 mm. It reads 3.15 mm for a wire. What is the true diameter?
- 3.17 mm
- 3.13 mm
- 3.15 mm
- 3.19 mm
Show answer
Answer: A. Corrected reading = observed − zero error = 3.15 − (−0.02) = 3.17 mm. A negative zero error is added.
-
4. A vernier caliper has a zero error of +0.05 cm. It reads 2.73 cm for an object. What is the true length?
- 2.78 cm
- 2.73 cm
- 2.68 cm
- 2.63 cm
Show answer
Answer: C. A positive zero error is subtracted: 2.73 − 0.05 = 2.68 cm.
-
5. Why is an instrument with a smaller least count preferred for the same measurement?
- It removes any zero error
- It makes the measured value larger
- It makes systematic errors impossible
- It gives a smaller percentage error
Show answer
Answer: D. The uncertainty of a reading is about one least count. A smaller least count is a smaller error on the same value, so the percentage error is smaller. Zero errors must still be checked and corrected.
Sign in to keep track of the lessons you finish.